EG560B4 Anesthesia Machine
EG560B4 Anesthesia Machine, with its ergonomic design, integrates a concise and smooth structure and high mobility. The microcomputer-controlled Ventilator adopts exported sensors and other components. Its scientific and proper design as well as rigorous production process are what demonstrate the inner quality of Aokai products: mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system. With multiple modern technologies applied, the Anesthesia Machine is both outstanding in performance and versatile in functions.

EG560B4 Anesthesia Machine, with its ergonomic design, integrates a concise and smooth structure and high mobility. The microcomputer-controlled Ventilator adopts exported sensors and other components. Its scientific and proper design as well as rigorous production process are what demonstrate the inner quality of Aokai products: mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system. With multiple modern technologies applied, the Anesthesia Machine is both outstanding in performance and versatile in functions.
 The Ventilator adopts microcomputer control with high-definition 5.7” LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
 The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose two kinds of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
 High-definition four-tube flowmeter, designed with stoppers for oxygen and nitrous oxide to control the output oxygen concentration.
 Reliable pneumatic transmission system.
 Economic and proper configuration of technologies, optional as required by the Customer.
 Modularized design and upgrading. Upgrading is available as required by the Customer.

Part of Anesthesia Machine
Main body                 High-strength engineering plastic rack, light,
beautiful and corrosion resistant
Scope of application          adult, child
Gas source                 O2: 0.27~0.55MPa    N2O: 0.27~0.55MPa
Flow meter                 O2:  0.1~1.0L/min     1.1~10L/min
N20: 0.1~1.0L/min     1.1~10L/min
O2, N2O linkage and N2O Stopper
  When using nitrous oxide, oxygen concentration>25%;
  When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply   25~75L/min
Low oxygen pressure alarm    There will be sound alarm when the oxygen
pressure < 0.2MPa
Vaporizer                  it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane,Isoflurane and Sevoflurane,two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
Respiratory circuit            working mode: all-close, semi-close, semi-open
                            APL: 0.5~7 kPa
Respiratory Bellows          bellows for adults, bellows for children
tidal volume range:0~1,500 ml

Part of Ventilator
Display mode               High-definition 5.7” LCD screen display
Ventilation mode            IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
Ventilation parameters        Tidal volume         50~1500 ml
Rate                2~99 bpm
SIMV rate           2~20 bpm
I:E                 2:1~1:8
Inspiratory trigger pressure  -1.0~2.0 kPa
PEEP                0~2.0 kPa
Pressure range        1.0~6.0 kPa
SIGH           1.5 times the inspiratory time
Parameters for ventilation monitoring  tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure
Monitoring of oxygen concentration  (21%~100%)
Security Alarm System
Oxygen Concentration alarm    upper limit setting range   (50%~100%)
low limit setting range    (15%~50%)
Airway pressure alarm         upper limit setting range   1.0~6.0 kPa
low limit setting range    0.4~2.0 kPa
Per-minute ventilation volume alarm   upper limit setting range   3.0~30L/min
low limit setting range    1.0~10L/min
Sustained high-pressure alarm       it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm             it will give sound and light alarm if there is
no tidal volume input for 15 seconds
Power alarm
 Power                       AC 220V±10%     50Hz±1Hz
                           (UPS, Storage battery)
(Mechanical Arms)
(End-expiratory CO2 monitoring)