某某清洁服务有限公司

SEARCH:
Content
EGM-2000AI (LCD) Electrosurgical Unit
EGM-2000AI (LCD) Electrosurgical Unit

Highest output: 400W
Nine working modes:
Monopolar:  pure cut,blend1, blend2, blend3, contact coagulation, forced coagulation, soft coagulation
Bipolar: bipolar coagulation, bipolar cut
The electrosurgical generator is used to control bleeding and surgical cutting in all general and open surgical procedures.
Liquid Crystal Display,with different audible and visual indication during the process of outputting.
Patient return electrode monitoring system and Power Peak System that provides surgeons with improved performance at Lower power settings, minimizing the risk of tissue damage.
Applicable for Prostate Vaporizing Surgical Operation(TUR), perfectly used under water. The electrosurgical generator can be cooperated with endoscope and laparoscope


Technical Specifications
Power: 220V±22V, 50Hz±1Hz(110V±11V,50/60Hz)
Operating frequency: a)Monopolar:512KHz  b) Bipolar:1024KHz
Power rating: 880VA±10%
Nine working modes:
Monopolar Cut
a) Pure cut:1W~400W( Load 800Ω)
b) Blend1:1W~300W(Load 800Ω)
c) Blend2:1W~200W(Load 800Ω)
d) Blend3:1W~150W(Load 800Ω)
Monopolar Coag
e) Contact Coag:1W~100W( Load 800Ω)
f) Forced Coag:1W~120W( Load 800Ω)
g) Soft Coag:1W~120W( Load 800Ω)
Bipolar
8) Bipolar Coagulation: 1W~50W( Load 100Ω)
9) Bipolar Cut: 1W~120W( Load 100Ω)
Power Consumption: ≤880VA