某某清洁服务有限公司

SEARCH:
Content
EG560B1 Anesthesia Machine (economy type)
EG560B1
Anesthesia Machine (economy type)

Part of Anesthesia
Main body                  High-strength engineering plastic rack, light,
beautiful and corrosion resistant
Scope of application           adult, child
Gas source                  O2, N2O: 0.27~0.55MPa
Flow meter                  O2: 0.1~1.0L/min    1.1~10L/min
Flow rate of rapid oxygen supply  25~75L/min
Vaporizer           Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer.
Respiratory circuit             working mode: all-close, semi-close, semi-open
                             releasing pressure:1~4 kPa
Respiratory bellows            bellows for adults, bellows for children
tidal volume range:0~1500 ml

Part of Ventilator
Display mode                 High-definition 4.8” LCD screen display
Ventilation  mode             IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate)
Ventilation parameters          Tidal volume         50~1500ml
Rate                4~80 bpm
I:E                2:1~1:4
Pressure range        1.0~6.0 kPa
Parameters for ventilation monitoring   tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform
Security Alarm System
Airway pressure alarm          upper limit setting range  1.0~6.0 kPa
low limit setting range    0.4~2.0 kPa
TV alarm                            upper limit setting range   0.5~2.0L
low limit setting range     0.02~1.0L
Sustained high-pressure alarm      it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm              it will give sound and light alarm if there is
no tidal volume input in 15s
Power alarm
Power                           AC 220V±10%     50Hz±1Hz

 

脚注栏目
自定内容
2006-2015 Everglorious Medical . All rights reserved